Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 21: CAL ADJ /5100 INSTRUMENT: Fluke 23: CAL ADJ /5100 INSTRUMENT: Fluke 73: CAL ADJ /5100 INSTRUMENT: Fluke 75: CAL ADJ /5100 INSTRUMENT: Fluke 77: CAL ADJ /5100 DATE: 15-Jul-92 AUTHOR: Fluke Corporation REVISION: $Revision: 2.3 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 36 CONFIGURATION: Fluke 5100B ============================================================================= # # Source: # Fluke 70 Series service manual (PN 731034). # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5100B accuracy file contains 6 month specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 6 month specifications of the 5100B are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- W 1.002 HEAD CALIBRATION ACCESS PROCEDURE 1.003 DISP Select the V DC function (auto range). 1.003 DISP 1.003 DISP Remove the four phillips-head screws from the bottom of 1.003 DISP the case. Turn the instrument face-up, and grasp the top 1.003 DISP cover at both sides. Pull the top cover from the UUT. 1.004 HEAD EQUIPMENT SETUP 1.005 DISP Ambient temperature: 18C - 28C. 1.005 DISP Relative humidity: less than 80%. 1.005 DISP Allow the UUT to stabilize. 1.006 HEAD CALIBRATION PROCEDURE 1.007 DISP Connect the 5100B and UUT as follows: 1.007 DISP [32] 5100B OUTPUT HI to UUT V Ohm Diode 1.007 DISP [32] 5100B OUTPUT LO to UUT COM 1.008 5100 3.000V S 2W 1.009 DISP Adjust the Calibration Pot for a reading of 3.000V. 1.012 END